The World Precision Instruments (WPI) EVOM2 is used for making trans epithelial electrical resistance (or TEER) measurements. Two silver/silver chloride pellets are used for the electrodes. The chloride ion tends to migrate, which is a reversible reaction. But, because of this chloride ion swapping, the electrodes need to be polarized before each use. Before using the meter for making voltage measurements, equilibrate the electrodes. This video shows you how.
- - April 23, 2013
A volt-ohm meter may apply a constant voltage of unknown current across the membrane and damage the cells electrically and also leave a chemical imbalance in the electrodes. WPI experimented with two volt-ohm meters:
- An expensive Fluke meter applies a DC voltage of 50 to 80mV, which (in a 1000Ω membrane) will pass 80µA. With a 200Ω membrane, this is 400µA.
- A discount meter puts out 500mV DC, which will pass 500µA through the membrane. With a 200Ω membrane, this is 2.5mA.