WPI’s EVOM™ Manual is the gold standard for delivering stable and repeatable Trans Epithelial Electrical Resistance (TEER) measurements. The EVOM™ Manual is the newest TEER meter. It qualitatively measures cell monolayer health and quantitatively measures cell confluence by determining an increase or a plateau in tissue resistance detected using our innovative EVOM™ technology. WPI’s state of the art EVOM™ technology provides you with real-time, valuable feedback during experimental measurements.
- - May 19, 2022
The new STX4 electrode works with the EVOM3, and the blades are replaceable. EVOM3 is used for making TEER (TER) Measurements in cell culture plates. Here's a quick tutorial to show you how to replace the blades.
- - May 05, 2022
The ATC2000 is a low noise heating system for maintaining animal body temperature during experimental procedures. It has three modes of operation, which we discuss here.
Normal mode compares the temperature and the setpoint for PID control. Adaptive mode is ingenious. It uses the monitored animal temperature and the actual plate temperature to define an offset value which is used to adjust the PID parameters. That offset is recalculated at a set interval for extremely precise control. Shutdown mode is a fail safe that shuts the unit down if the plate temperature reaches a defined setpoint (like 45ºC) that would endanger the animal.
The instrument has three modes of operation:
- Normal
- Adaptive
- Shutdown
Normal Mode
The normal mode compares the temperature of the configured sensor (RTD or TC) with the set point to control the heater.
Adaptive Mode
In the adaptive mode, the sensor in the heated plate is used as the control trigger for the set point. A platinum sensor in the plate monitors...more
- - May 04, 2022
Sapphire Blades are a precise and durable solution for use in microsurgery, dissection, and related applications.
How are Sapphire Blades Manufactured?
Sapphire blades are created using a chemical process which forms an artificial sapphire. The sapphire has a hardness value of 9 on the Mohs hardness scale, slightly below the hardest mineral on earth- the diamond. This characteristic provides supreme durability and sharpness to the blade.
Why Choose a Sapphire Blade?
...moreUltra Sharp
The sapphire blade's ultra sharp edge creates fine and precise cuts that prevent tissue damage and promote faster healing with less scarring.
Easier Use
The sapphire knife's ergonomic grip offers an easier grasp and is designed to be used with WPI's blades.
Longer Lasting
The blade's chemical composition provides a durable edge that lasts longer than steel surgical instruments, saving costs for your practice.
Greater Precision
The blade's sharp point and transparent finish allows greater precision in hair restoration,
- - May 03, 2022
The ATC2000 is a low noise heating system for maintaining animal body temperature during experimental procedures. Here we show you how to reset the factory defaults on your ATC2000.
To reset all the configuration parameters to the factory defaults:
- Turn the unit off.
- Then, hold down both the Up and Down buttons while you turn the unit on again.
All customized paramaters will be set to the original factory default settings.
If you have any questions, just give us a call.
- - May 01, 2022
The ATC2000 is a low noise heating system for maintaining animal body temperature during expiremental procedures.
Here are some common issues you may encounter:
- No power. If you have no power, you may have a poor connection. Check the power cable connection.
- ERROR displays instead of a temperature. If ERROR displays instead of a temperature on the main display, you may have a poor probe connection. Disconnect the probe and properly re-connect it to the ATC2000.
- Animal cannot reach set temperature. If the animal cannot reach the set temperature, you may have insufficient heating power. The maximum power output on the heating device is 30W. In most situations the power is sufficient to keep the animal warm up to 40˚ C. To minimize heating loss, insulate the bottom of heating plate and cover the animal with a small blanket to keep it warm.
- Alarm is sounding. If an alarm is sounding, you may have poor transfer of heat from the heating plate to the animal. (The plate is hot, but the animal...more
- - April 28, 2022
The ATC-2000 is a low noise heating system for maintaining animal body temperature during experimental procedures. The ATC-2000 uses a digital PID controller to regulate the application of power to the heating plate to achieve the desired temperature at the monitored sensor. PID stands for Proportional, Integral and Differential. If the factory default PID settings do not adequately stabilize the system, you can manually adjust the PID parameters. WPI doesn't reccomend changing these parameters manually unless you are familiar with the functions of each PID parameter.
Setting PID Parameters
- Press the PID button to display the PID screen. The two arrows on the screen indicate which digit is being adjusted.
2. Press the Config button to select the P line to modify the proportional value, I line to modify the integral value or the D line to modify the differential value.
3. Press the PID button to select the digit to modify. Each time you press the PID button, the top arrow moves...more - - April 27, 2022
The ATC2000 is a low noise heating system for maintaining animal body temperature during experimental procedures. The ATC2000 uses a digital PID controller to regulate the application of power to the heating plate to achieve the desired temperature at the monitored sensor. PID stands for Proportional, Integral and Differential. Here’s how PID control works:
The default set point is 37°C, but you can configure it. The controller calculates the error, which is the difference between the present temperature and the desired temperature.
Error = Monitored temperature – Set point
Three individual components are derived by applying different functions to the error.
The default set point is 37°C, but you can configure it. The controller calculates the error, which is the difference between the present temperature and the desired temperature.
Error = Monitored temperature – Set point
Three individual components are derived by applying different functions to the
- The error is multiplied by the proportional...more
- - April 26, 2022
ATC2000 is a low noise heating system for maintaining animal body temperature during experimental procedures. Here we will look at how to adjust the basic configuration parameters. From the configuration menu, you can:
- Adjust the set point temperature
- Set the high and low alarm limits
- Enable/disable the audible alarm
- Adjust the three temperature offsets
- Enable adaptive mode and setup the adaptive interval
- Change the shutdown temperature setpoint
You can use the same procedure to adjust any ATC2000 configuration parameter, regardless of which parameter you’re adjusting.
- Press the Config button to access the Configuration menu. Press the Config button again to toggle to the parameter you want to adjust.
- Then use the Up or Down button to make an adjustment.
- Press the Display button to save the parameter.
Temperature Setpoint
You may adjuSetting Low Alarm Limit
Likewise, you may set the Low Alarm Limit. If the controlled temperature falls below the low alarm temperature...more